Multi Gas Analyzer MGA O2/CO2/Ethyleen dual range analyzer

Multi Gas Analyzer MGA O2/CO2/Ethyleen dual range analyzer

The Storex MGA O2/CO2/Ethylene Dual range analyzer was developed by Storex.

Key features:

  • Hardware platform for 4 analyzers with choice between:
    • O2 chemical cell
    • CO2 Infrared Carbocap technology (standard CO2 analyzer in Storex O2/CO2 analyzers)
    • Choice of 1, or 2 ethylene sensors
    • NH3 sensor
    • Or any other type of sensor according to customer's requirements

 

  • Large backlit display
  • Oxygen sensor lifetime indicator.
  • Power and battery indicator.
  • Lithium Ion battery pack rechargeable (no more Ni Cad batteries).
  • Easy to operate and calibrate.


Output : O2 / CO2 4-20 mA / 0-20 mA

  • Output                     : O2 / CO2 / NH3 / C2H4 Modbus
  • O2 sensor               : Chemical oxygen cell
  • Range                     : 0-25% O2
  • Accuracy 0-3%       : 0.01% O2
  • Accuracy 3-25%     : 0.1% O2
  • O2 cell life               : 2 years with regular use.
  • CO2 sensor            : Carbocap CO2 meter type GMP 251
  • Range                     : 0-20% CO2
  • Compensation    Pressure and temperature compensation integrated.
  • Compensation for background gases
  • Head of sensor is heated to prevent condensation.
  • Temperature range : -40 - +60 ⁰C
  • Enclosure                : IP65

The MGA analyzer is designed for application in central measuring systems for CA/ULO cells with built-in ethylene analyzer based on:

  •  2 ethylene sensors are incorporated:
    • 1 sensor with a range of 0-20 ppm ethylene, resolution 0.1 ppm accuracy approx. +/- 0.2 ppm
    • 1 sensor with a range of 0-200 ppm ethylene resolution 1 ppm, accuracy up to 100 ppm approx. 2-3 ppm, above 100 ppm approx. 5% of the measured value.

The MGA analyzer for installation in measurement system or as a portable analyzer:

The MGA analyzer that we now have available was developed for installation in our Storex measurement systems / CO2 scrubbers for automatic measurement together with the O2/CO2 content. The development of the ethylene content can then also be displayed in graphs. 

Technical details:

Housing           : Aluminum housing with large display,

Dimensions      : 23.5 X 25.5 X 14 CM

Weight             : 3,2 kg.

Components    : Pump, flow sensor, operating program, inlet/outlet for gas

Attachments    : Strap, tube + filter and handle for carrying

Flow rate          : 0.3 L/Minute

Voltage            : 24VDC

Battery             : Rechargeable lithium ion battery